Neuroscan SynAmp2

장비기능 : 두피에 전극을 부착하여 사용자의 뇌파 신호 습득
주요용도 : 뇌-컴퓨터 인터페이스를 위한 동작 상상 및 시각 자극에 대한 뇌파 습득 및 분석
주요사양 :
32-채널 EEG 전극 및 3개의 EOG 전극
Monopolar Channel Count......64 Per Headbox, 256 Per System Unit
Bipolar Channel Count...........4 Per Headbox, 16 Per System Unit
High Level Channel Count.....2 Per Headbox, 8 Per System Unit
Maximum Sampling Rate.......20,000 Hz / Channel all channels driven
Sampling Method..................All Channels Sampled Simultaneous
A/D Resolution.....................24 Bit
Input Impedance..................10 GOhms
CMRR...................................110 dB
Input Noise...........................<0.5 uV RMS DC to 200Hz,<1.5uV RMS
Bandwidth...........................DC-3500 Hz
Low Pass Filter (Maximum)..-3 dB at 3500 Hz, single pole
High Pass Filter (Minimum)...DC or -3 dB at .05 Hz, single pole
Full Scale Input Range (DC Mode).....+/- 200 mV
System Gain (DC Mode)......10
Sensitivity (DC Mode)..........24 nV/bit
Full Scale Input Range (AC Mode).....+/- 995uV
System Gain (AC Mode)......2010
Sensitivity (AC Mode)..........3 nV/bit
Digital (TTL) Inputs...............8 Bit Stimulus, 8 Bit Response
High Level Input Range........+/- 5 V
Impedance Measurement.....1K Ohm to 50 K Ohm
IMRR....................................127 dB
Headbox Size (h x w x d)...4.5 x18.7 x 21.8cm, per 64 channels
Headbox Weight..................1.5 kg
System Unit Size.................23 x 11.6 x 32.5 cm per 256 channels
System Unit Weight.............4.3 kg
Power Unit Size..................23 x 11.6 x 32.5 cm per 256 channels
Power Unit Weight..............12.5 kg
Device Certification
EU Risk Classification 2B
Type CF
FDA Class II
FDA Product Code GWQ, GWP, GWF, GWE, GWJ
Safety Specifications
EN 60601-1 Medical Electrical Devices - General Requirements for Safety
UL2601.1 - Including US National Deviations for IEC601-1
CSA22.2 - Including Canadian National Deviations for IEC601-1
EN 60601-1-1 - Collateral Standard - Systems
EN 60601-1-2 - Collateral Standard - Electromagnetic Compatibility
EN 60601-1-4 - Collateral Standard - Programmable Systems
EN 60601-2-26 - Specific requirements for electroencephalographs